IEC 61010-1:2010/AMD1:2016/COR1:2019

Corrigendum 1 - Amendment 1 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements

IEC 61010-1:2010/AMD1:2016/COR1:2019

Name:IEC 61010-1:2010/AMD1:2016/COR1:2019   Standard name:Corrigendum 1 - Amendment 1 - Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 1: General requirements
Standard number:IEC 61010-1:2010/AMD1:2016/COR1:2019   language:English and French language
Release Date:20-Mar-2019   technical committee:TC 66 - Safety of measuring, control and laboratory equipment
Drafting committee:WG 1 - TC 66/WG 1   ICS number:19.080 - Electrical and electronic testing

 IEC 2019
INTERNATIONAL ELECTROTECHNICAL COMMISSION
COMMISSION ÉLECTROTECHNIQUE INTERNATIONALE
____________
IEC 61010-1:2010/AMD1:2016 IEC 61010-1:2010/AMD1:2016
Edition 3.0  2010-06 Édition 3.0  2010-06

Amendment 1:2016 Amendement 1:2016

Safety requirements for electrical equipment for Règles de sécurité pour appareils électriques de
measurement, control, and laboratory use – mesurage, de régulation et de laboratoire –

Part 1: General requirements Partie 1: Exigences générales

CO RRI G E NDUM 1
Corrections to the French version appear after the English text.
Les corrections à la version française sont données après le texte anglais.

Replace Table I.1 with the following:
Table I.1 – Line-to-neutral voltages for common MAINS supply systems
– 2 – IEC 61010-1:2010/AMD1:2016/COR1:2019
 IEC 2019
MAINS systems and nominal voltages Line-to-
neutral
Three-phase Three-phase Three-phase Three-phase Single-phase Single-
voltage
four-wire four-wire three-wire three-wire two-wire phase (split-
pertinent
a a
systems systems systems phase)
systems systems
to MAINS
three-wire
system
a
unearthed with earthed a.c. or d.c.
with earthed with systems
type and
phase
neutral unearthed a.c. or d.c.
nominal
neutral (IT
voltage
TT system
b, c
systems)
TN-C-S System
V V V V V V V
12,5 to 48 30/60 50
66/115 66 60 100
d
120/208 120/208 110, 115 100 100 100/200 150
110, 115
127/220 120, 127 120 110/220
120, 127
115/230
120/240
220/380 230/400 200 200 220 220/440 300
230/400 277/480 220, 230, 240 240 230 240/480
240/415 260, 277, 347 240
260/440 380, 400, 415
277/480 440, 480
347/600 347/600 500 347 480 480/960 600
380/660 400/690 577 380, 400, 415
400/690 600 440, 480, 600
417/720
480/830
660 1 000 1 000
690, 720
830, 1 000
a
Voltages shown as two voltages separated by a “/” represent the phase-to-neutral (or line-to-neutral) voltage
followed by the phase-to-phase (or line-to-line) voltage. For example, “120/208” indicates that the voltage from
any phase to neutral is 120 V, and the voltage from any phase to another phase is 208 V. Likewise, “220/440”
indicates that the voltage from either line-to-neutral
...

  • Relates Information
  • IEC 60300-3-7:1999

    IEC 60300-3-7:1999 - Dependability management - Part 3-7: Application guide - Reliability stress screening of electronic hardware Released:5/31/1999 Isbn:2831847974
    09-20
  • HD 571 S1:1990

    HD 571 S1:1998
    09-20
  • ISO 8130-4:1992

    ISO 8130-4:1992 - Coating powders
    09-20
  • HD 478.2.7 S1:1990

    HD 478.2.7 S1:2003
    09-19
  • ISO 8473:1988/Cor 1:1992

    ISO 8473:1988/Cor 1:1992 - Information processing systems — Data communications — Protocol for providing the connectionless-mode network service — Technical Corrigendum 1 Released:12/10/1992
    09-19
  • EN ISO 9013:2017/prA1

    EN ISO 9013:2017/oprA1:2024
    09-19
  • IEC 60118-6:1999

    IEC 60118-6:1999 - Hearing aids - Part 6: Characteristics of electrical input circuits for hearing aids Released:6/9/1999 Isbn:2831848075
    09-19
  • HD 280.3 S1:1990

    HD 280.3 S1:1999
    09-19
  • ISO 9832:1992

    ISO 9832:1992 - Animal and vegetable fats and oils -- Determination of residual technical hexane content
    09-19
  • EN 60188:1988/A1:1990

    EN 60188:1999/A1:1999
    09-18