SLOVENSKI STANDARD
01-julij-2021
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-091. del: Posebne zahteve za rentgenske sisteme v ohišjih
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-091: Particular requirements for cabinet X-ray systems
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2
-091: Besondere Anforderungen für Röntgengeräteschränke
Exigences de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-091: Exigences particulières pour les équipements à rayons X
montés en armoire
Ta slovenski standard je istoveten z: EN IEC 61010-2-091:2021/A11:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN IEC 61010-2-
091:2021/A11
NORME EUROPÉENNE
EUROPÄISCHE NORM
April 2021
ICS 19.080; 71.040.10
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-091: Particular requirements
for cabinet X-ray systems
Règles de sécurité pour appareils électriques de mesurage, Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
de régulation et de laboratoire - Partie 2-091: Exigences Regel- und Laborgeräte - Teil 2-091: Besondere
particulières pour les équipements à rayons X montés en Anforderungen für Röntgengeräteschränke
armoire
This amendment A11 modifies the European Standard EN IEC 61010-2-091:2021; it was approved by CENELEC on 2021-04-01.
CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
amendment the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 61010-2-091:2021/A11:2021 E
European foreword
This document (EN IEC 61010-2-091:2021/A11:2021) has been prepared by CLC/TC 66X “Safety of
measuring, control, and laboratory equipment”.
The following dates are fixed:
• latest date by which this document has (dop) 2022-04-01
to be implemented at national level by
publication of an identical national
standard or by endorsement
• latest date by which the national (dow) 2024-04-01
standards conflicting with this document
have to be withdrawn
This document amends EN IEC 61010-2-091:2021.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a mandate given to CENELEC by the European Commission
and the European Free Trade Association, and supports essential requirements of EU Directive(s).
For the relationship with EU Directive(s) see informative Annex ZZ, which is an integral part of this
document.
1 Modification to 15.1, “General”
Replace 15.1 with the following:
“Replace the existing text and conformity statement with the following new text:
INTERLOCKs used to protect OPERATORS or bystanders both against X-radiation HAZARDS and
other HAZARDS shall meet the requirements of 15.2, 15.3, and 15.101.
INTERLOCKs used only to protect OPERATORS or bystanders against X-radiation HAZARDS shall
meet the requirements of 15.101.
INTERLOCKs used only to protect OPERATORS or bystanders against other HAZARDS shall meet
the requirements of 15.2 and 15.3.
Conformity is checked by inspection and as specified in 15.2, 15.3, and 15.101, as applicable.”
2 Addition of Annex ZA, "Normative references to international publications
with their corresponding European publications"
Add the following Annex ZA:
"
Annex ZA
(normative)
Normative references to international publications
with their corresponding European publications
The following documents, in whole or in part, are normatively referenced in this do
...