SLOVENSKI STANDARD
01-december-2021
Varnostne zahteve za električno opremo za meritve, nadzor in laboratorijsko
uporabo - 2-061. del: Posebne zahteve za laboratorijske atomske spektrometre s
termično atomizacijo in ionizacijo
Safety requirements for electrical equipment for measurement, control and laboratory
use - Part 2-061: Particular requirements for laboratory atomic spectrometers with
thermal atomization and ionization
Sicherheitsbestimmungen für elektrische Mess-, Steuer-, Regel- und Laborgeräte - Teil 2
-061: Besondere Anforderungen an Labor-Atomspektrometer mit thermischer
Atomisierung und Ionisation
Exigences de sécurité pour appareils électriques de mesurage, de régulation et de
laboratoire - Partie 2-061: Exigences particulières pour spectromètres atomiques de
laboratoire avec vaporisation et ionisation thermiques
Ta slovenski standard je istoveten z: EN IEC 61010-2-061:2021/A11:2021
ICS:
19.080 Električno in elektronsko Electrical and electronic
preskušanje testing
71.040.10 Kemijski laboratoriji. Chemical laboratories.
Laboratorijska oprema Laboratory equipment
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.
EUROPEAN STANDARD EN IEC 61010-2-
061:2021/A11
NORME EUROPÉENNE
EUROPÄISCHE NORM
September 2021
ICS 19.080; 71.040.20
English Version
Safety requirements for electrical equipment for measurement,
control and laboratory use - Part 2-061: Particular requirements
for laboratory atomic spectrometers with thermal atomization and
ionization
Exigences de sécurité pour appareils électriques de Sicherheitsbestimmungen für elektrische Mess-, Steuer-,
mesurage, de régulation et de laboratoire - Partie 2-061: Regel- und Laborgeräte - Teil 2-061: Besondere
Exigences particulières pour spectromètres atomiques de Anforderungen an Labor-Atomspektrometer mit thermischer
laboratoire avec vaporisation et ionisation thermiques Atomisierung und Ionisation
This amendment A11 modifies the European Standard EN IEC 61010-2-061:2021; it was approved by CENELEC on 2021-08-02.
CENELEC members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this
amendment the status of a national standard without any alteration.
Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN-CENELEC
Management Centre or to any CENELEC member.
This amendment exists in three official versions (English, French, German). A version in any other language made by translation under the
responsibility of a CENELEC member into its own language and notified to the CEN-CENELEC Management Centre has the same status as
the official versions.
CENELEC members are the national electrotechnical committees of Austria, Belgium, Bulgaria, Croatia, Cyprus, the Czech Republic,
Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, the
Netherlands, Norway, Poland, Portugal, Republic of North Macedonia, Romania, Serbia, Slovakia, Slovenia, Spain, Sweden, Switzerland,
Turkey and the United Kingdom.
European Committee for Electrotechnical Standardization
Comité Européen de Normalisation Electrotechnique
Europäisches Komitee für Elektrotechnische Normung
CEN-CENELEC Management Centre: Rue de la Science 23, B-1040 Brussels
© 2021 CENELEC All rights of exploitation in any form and by any means reserved worldwide for CENELEC Members.
Ref. No. EN IEC 61010-2-061:2021/A11:2021 E
European foreword
This document (EN IEC 61010-2-061:2021/A11:2021) has been prepared by CLC/TC 66X “Safety of
measuring, control, and laboratory equipment”.
The following dates are fixed:
• latest date by which this document has to be (dop) 2022-08-02
implemented at national level by publication of
an identical national standard or by
endorsement
• latest date by which the national standards (dow) 2024-08-02
conflicting with this document have to be
withdrawn
This document amends EN IEC 61010-2-061:2021.
Attention is drawn to the possibility that some of the elements of this document may be the subject of
patent rights. CENELEC shall not be held responsible for identifying any or all such patent rights.
This document has been prepared under a Standardization Request given to CENELEC by the
European Commission and the European Free Trade Association, and supports essential requirements
of EU Directive(s) / Regulation(s).
For relationship with EU Directive(s) / Regulation(s), see informative Annex ZZ, which is an integral part
of this document.
Any feedback and questions on this document should be directed to the users’ national committee. A
complete listing of these bodies can be found on the CENELEC website.
1 Modification to Clause 2 “Normative references”
Add the following normative references:
“IEC 62061, Safety of machinery — Functional safety of safety-related electrical, electronic and
programmable electronic control systems
ISO 13849-1, Safety of machinery — Safety-related parts of control systems — Part 1: General
principles for design
ISO 13849-2, Safety of machinery — Safety-related parts of control systems — Part 2: Validation"
2 Modification to 4.4.2.101, “Sampling probe tip”
Replace the entire subclause by the following:
“Any system designed to withdraw a sampling probe tip after sampling has been completed shall be
overridden, so as to leave the tip in its most exposed position when a sample vessel is removed.
Exception:
The withdrawal system need not be overridden if the sampling probe is designed in such a manner that
no SINGLE FAULT CONDITION can cause the tip to remain exposed after sampling has been completed.”
3 Modification to 5.1.5.101, “Gas and liquid connections”
Replace the entire subclause by the following:
“The following shall be unambiguously marked adjacent to the connector on the equipment (see 5.2):
a) the identity of the gas or liquid;
b) the maximum permitted pressure,
c) flow direction of the gas and liquid, if applicable.
NOTE Such markings can be specific (for example acetylene, propane, water) or generic (for example fuel
gas, oxidant gas, coolant, waste liquid).
Where no internationally recognized symbol (such as a chemical formula) exists, the equipment shall
be marked with symbol 14 of Table 1.
Conformity is checked by inspection.”
4 Modification to 5.4.3, “Equipment installation”
Replace list item bb) by the following:
“bb) requirements for a fume extraction system to remove exhaust gases which could be hazardous
(see 13.1);”
5 Modification to 8.1, “General”
Replace the text of item 3) by the following text:
“3) except for FIXED EQUIPMENT and equipment with a mass over 100 kg, the test of 8.3.1 or 8.3.2, as
applicable. The equipment is not operated during the tests.”
6 Modification to 10.1, “Surface temperature limits for protection against
burns”
Replace the newly added paragraph by the following text:
“Protection shall be provided to prevent unintentional direct access to hot gases or plasma emerging
from the equipment by
...