SIST EN 12470-1:2000+A1:2009

Clinical thermometers - Part 1: Metallic liquid-in-glass thermometers with maximum device

SIST EN 12470-1:2000+A1:2009

Name:SIST EN 12470-1:2000+A1:2009   Standard name:Clinical thermometers - Part 1: Metallic liquid-in-glass thermometers with maximum device
Standard number:SIST EN 12470-1:2000+A1:2009   language:English language
Release Date:09-Sep-2009   technical committee:VAZ - Healthcare
Drafting committee:   ICS number:17.200.20 - Temperature-measuring instruments
2003-01.Slovenski inštitut za standardizacijo. Razmnoževanje celote ali delov tega standarda ni dovoljeno.Medizinische Thermometer - Teil 1: Mit metallischer Flüssigkeit gefüllte Glasthermometer mit MaximumvorrichtungThermomètres médicaux - Partie 1: Thermomètres à dilatation de liquide métallique dans une gaine de verre, avec dispositif à maximumClinical thermometers - Part 1: Metallic liquid-in-glass thermometers with maximum device17.200.20Instrumenti za merjenje temperatureTemperature-measuring instrumentsICS:Ta slovenski standard je istoveten z:EN 12470-1:2000+A1:2009SIST EN 12470-1:2000+A1:2009en,fr,de01-oktober-2009SIST EN 12470-1:2000+A1:2009SLOVENSKI
STANDARD



SIST EN 12470-1:2000+A1:2009



EUROPEAN STANDARD NORME EUROPÉENNE EUROPÄISCHE NORM
EN 12470-1:2000+A1
June 2009 ICS 17.200.20 Supersedes EN 12470-1:2000English Version
Clinical thermometers - Part 1: Metallic liquid-in-glass thermometers with maximum device
Thermomètres médicaux - Partie 1: Thermomètres à dilatation de liquide métallique dans une gaine de verre, avec dispositif à maximum
Medizinische Thermometer - Teil 1: Mit metallischer Flüssigkeit gefüllte Glasthermometer mit Maximumvorrichtung This European Standard was approved by CEN on 13 May 1999 and includes Amendment 1 approved by CEN on 16 May 2009.
CEN members are bound to comply with the CEN/CENELEC Internal Regulations which stipulate the conditions for giving this European Standard the status of a national standard without any alteration. Up-to-date lists and bibliographical references concerning such national standards may be obtained on application to the CEN Management Centre or to any CEN member.
This European Standard exists in three official versions (English, French, German). A version in any other language made by translation under the responsibility of a CEN member into its own language and notified to the CEN Management Centre has the same status as the official versions.
CEN members are the national standards bodies of Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom.
EUROPEAN COMMITTEE FOR STANDARDIZATION
COMITÉ EUROPÉEN DE NORMALISATION EUROPÄISCHES KOMITEE FÜR NORMUNG
Management Centre:
Avenue Marnix 17,
B-1000 Brussels © 2009 CEN All rights of exploitation in any form and by any means reserved worldwide for CEN national Members. Ref. No. EN 12470-1:2000+A1:2009: ESIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 2 Contents Page Foreword .31Scope .42Normative references .43Definitions .44Unit .55Types of thermometer .56Requirements .57Test methods .88Information supplied by the manufacturer . 14Annex A (informative)
Suggested types of testing for the requirements of this standard . 16Annex B (informative)
Advice to be considered for inclusion in the instruction leaflet accompanying mercury-in-glass thermometers . 18Annex C (informative)
Bibliography . 19Annex ZA (informative)
!!!!Relationship between this European Standard and the Essential Requirements of EU Directive 93/42/EEC"""" . 20Annex ZB (informative)
A-deviations . 23 SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 3 Foreword This document (EN 12470-1:2000+A1:2009) has been prepared by Technical Committee CEN/TC 205 "Non-active medical devices", the secretariat of which is held by DIN. This European Standard shall be given the status of a national standard, either by publication of an identical text or by endorsement, at the latest by December 2009, and conflicting national standards shall be withdrawn at the latest by March 2010. This document includes Amendment 1, approved by CEN on 2009-05-16. This document supersedes EN 12470-1:2000. The start and finish of text introduced or altered by amendment is indicated in the text by tags ! ". This European Standard has been prepared under a mandate given to CEN by the European Commission and the European Free Trade Association, and supports essential requirements of EU Directive(s). For relationship with EU Directive(s), see informative Annex ZA, which is an integral part of this standard.
For A-deviations, see annex ZB. This European Standard applies to clinical thermometers which are used for measuring the body temperature of humans. EN 12470 consists of the following Parts under the general title 'Clinical thermometers': Part 1: Metallic liquid-in-glass thermometers with maximum device Part 2: Phase change-type (dot matrix) thermometers Part 3: Performance of compact electrical thermometers (non-predictive and predictive) with maximum device Part 4: Performance of electrical thermometers for continuous measurements Part 5: Performance of infra-red ear thermometers (with maximum device) Annexes A, B, C, ZA and ZB are informative. According to the CEN/CENELEC Internal Regulations, the national standards organizations of the following countries are bound to implement this European Standard: Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Latvia, Lithuania, Luxembourg, Malta, Netherlands, Norway, Poland, Portugal, Romania, Slovakia, Slovenia, Spain, Sweden, Switzerland and United Kingdom. SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 4 1 Scope This Part of EN 12470 specifies performance requirements and test methods for clinical liquid-in-glass thermometers with maximum device and applies only to thermometers filled with metallic liquid. NOTE 1 Note that in some European countries the use of mercury is prohibited in clinical thermometers.
NOTE 2 Substances other than metallic liquids can be used in the manufacturing of liquid-in-glass thermometers. No reference is made to these in this European standard because there is no experience of clinical thermometers which use other substances. This European Standard does not apply to clinical thermometers designed for special applications (e.g. thermometers for premature babies, ovulation thermometers) which, owing to their measurement range, scale interval or maximum permissible error, fall outside the scope of this standard. 2 Normative references This European Standard incorporates by dated or undated reference, provisions from other publication. These normative references are cited at the appropriate places in the text and the publications are listed hereafter. For dated references, subsequent amendments to or revisions of any of these publications apply to this European Standard only when incorporated in it by amendment or revision. For undated references the latest edition of the publication referred to applies. EN 980, Graphical symbols for use in the labelling of medical devices EN 1041, Information supplied by the manufacturer with medical devices ISO 719, Glass - Hydrolytic resistance of glass grains at 98 degrees C - Method of test and classification ISO 2859-2:1985, Sampling procedures for inspection by attributes – Part 2: Sampling plans indexed by limiting quality (LQ) for isolated lot inspection 3 Definitions For the purposes of this Part of EN 12470, the following definitions apply: 3.1 correction value added algebraically to the uncorrected result of a measurement to compensate for systematic error 3.2 error result of measurement minus a true value of the measurand 3.3 maximum device device which prevents the liquid column from falling when the temperature of the liquid in the bulb returns to the ambient temperature 3.4 scale panel (enclosed-scale type) panel to which the scale is fixed longitudinally behind the capillary tube 3.5 stabilized thermometer reading SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 5 result given by a thermometer which, after attaining thermal equilibrium with a water bath of a temperature within the thermometers measuring range, has been removed from the water bath and cooled to a temperature of between 15 ûC and 30 ûC 3.6 zero point correction correction of the reading of the thermometer at 0 ûC NOTE K = 0 ûC -t1 3.7 zero point depression change of zero point correction after heating and fast cooling of the thermometer 4 Unit The unit of temperature shall be the degree Celsius, symbol ûC. 5 Types of thermometer The types of metallic liquid-in-glass thermometers with maximum device shall be: a) solid-stem, or b) enclosed-scale 6 Requirements 6.1 Scale 6.1.1 Measuring range and scale interval The thermometer shall cover the minimum range from 35,5 ûC to 42,0 ûC with a scale interval of 0,1 ûC. Testing shall be performed by visual inspection. 6.1.2 Scale marks and numbering 6.1.2.1 General The scale marks shall be uniformly spaced and of uniform width.
The scale marks and numbers shall be at right angles to the axis of the thermometer and shall be visible at the same time as the liquid column. Testing shall be performed by visual inspection. 6.1.2.2 Scale spacing The distance between adjacent scale marks shall be at least 0,5 mm for solid-stem thermometers and at least 0,6 mm for enclosed-scale thermometers. Testing shall be performed using an appropriate device, which magnifies the scale by at least x 4. SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 6
6.1.2.3 Width and length of the scale marks The scale marks shall be durably marked and shall be of uniform width not exceeding one-quarter of the length of a scale spacing plus 0,05 mm for solid-stem thermometers, or one-fifth of the length of a scale spacing plus 0,05 mm for enclosed-scale thermometers. Scale marks representing whole degrees and half degrees shall be longer than the other scale marks. Testing shall be performed using an appropriate device, which magnifies the scale by at least x 4. 6.1.2.4 Numbering Scale marks representing whole degrees shall be numbered. NOTE The scale mark representing the temperature of 37 ûC can be rendered conspicuous by means of a different colour from that used for the numbering and/or by additional marking. Testing shall be performed by visual inspection. 6.1.2.5 Special requirements for solid-stem thermometers The scale shall be indelibly marked directly on the thick-walled capillary stem. When tested in accordance with 7.2, the appearance of the scale lines shall not be significantly affected. 6.1.2.6 Special requirements for enclosed-scale thermometers The thermometer shall have a separate scale panel adjacent to the capillary tube. The capillary tube and the scale panel shall be enclosed in a transparent tube impermeably fixed to the bulb and forming a protection sheath. The scale shall be marked on a scale panel fixed longitudinally behind the capillary tube. Testing shall be performed by visual inspection. 6.1.2.7 Resistance to breakage of enclosed-scale thermometers The thermometer shall not break when subjected to a force of at least 50N in accordance with 7.8. 6.2 Material 6.2.1 Thermometer bulb 6.2.1.1 General The thermometer bulb shall be made of a type of glass which satisfies the requirements specified in 6.2.1.2 and 6.2.1.3. This glass shall be clearly and indelibly identified by either: a) an integral mark introduced by the glass manufacturer in such a way as to be clearly recognizable on the bulb after manufacture of the thermometer or; b) by a mark chosen by the glass manufacturer and affixed by the thermometer manufacturer and clearly indicating the type of glass used. SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 7 6.2.1.2 Hydrolytic resistance When tested in accordance with ISO 719 the quantity of alkali obtained in solution from 1 g of the glass shall not exceed 263,5 µg of Na2O. 6.2.1.3 Zero point depression When tested in accordance with the method given in 7.3, the glass shall have an average zero point depression not exceeding 0,05 °C. 6.2.2 Maximum device and capillary The glass used for the maximum device and the capillary shall have the same hydrolytic resistance as the glass for the thermometer bulb. 6.2.3 Scale panel of enclosed-scale thermometer The scale panel of enclosed-scale thermometers shall be of opal glass, metal or another material of a dimensional stability such that, when tested in accordance with 7.4, the length of the scale sample after heating shall not differ from that before heating by more than 0,2 %. 6.2.4 Thermometric liquid filling The metallic liquid filling used in the thermometer shall have the purity, properties and characteristics that will enable the finished thermometers to comply with all requirements specified in this European Standard. The bulb, the capillary tube and the liquid column of the finished thermometer shall be free of gas, water, glass fragments or other foreign material. Testing shall be performed by visual inspection using an appropriate device which magnifies by at least x 4. 6.3 Manufacture 6.3.1 Freedom from defects The thermometer and its surface shall be free from defects that are likely to interfere with its proper functioning or to mislead the user, e.g. errors in graduation, flaws in the glass and constructional defects. Testing shall be performed with an appropriate device, which magnifies the scale by at least x 4.
6.3.2 Capillary tube The capillary tube shall ensure that the entire length of the liquid column and the meniscus are clearly visible from at least one angle. It shall be prismatic in form and have a magnifying effect, or shall be so designed as to ensure consistent ease of reading. Testing shall be performed by visual inspection. 6.3.3 Thermometer ends The ends of the thermometer shall be smoothly rounded in order to prevent tissue damage during use.
Testing shall be performed by visual and tactile inspection. SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 8 6.3.4 Liquid column When the thermometer is heated slowly, the liquid column shall rise uniformly without appreciable surges. NOTE Surges larger than half a scale interval are considered as defects. Testing shall be performed by visual inspection. When tested in accordance with 7.6, the liquid column shall fall below the lowest numbered scale mark. The thermometer shall not break when subjected to this test. 6.3.5 Scale panel of enclosed-scale thermometer The scale panel shall be either directly fixed to the capillary tube or be in direct contact with the capillary tube and fixed sufficiently firmly in the sheath to prevent any displacement with respect to the capillary tube. The scale panel shall be so positioned as to ensure that any displacement can be easily detected either by means of an indelible mark on the sheath level with a numbered scale mark or by some equivalent means. The marking shall be in accordance with 6.1.2.3 and shall be placed at 38 ûC with a lateral tolerance of one-fifth of a scale spacing. The sheaths of enclosed-scale thermometers shall be free of moisture, liquid, glass fragments and foreign material. Testing shall be performed by visual inspection. 6.4 Effect of storage When tested in accordance with 7.7, the thermometer shall meet the requirements specified in 6.5. 6.5 Maximum permissible error under reference conditions Within the measuring range specified in 6.1.1 the maximum permissible error of the thermometer shall be + 0,1 ûC and - 0,15 ûC. This value applies to the stabilized thermometer readings. Test shall be performed in accordance with 7.9. 6.6 Influence of immersion time The thermometer at an ambient temperature t1 of (23 ± 2) ûC is rapidly immersed for 20 s in a vigorously stirred water bath at a constant temperature t2 of (38 ± 2) ûC, withdrawn and read after at least 1 min. Then the procedure is repeated with an immersion time of 60 s. The readings for 60 s and 20 s shall not differ more than 0,006·(t2 – t1). Testing shall be performed in accordance with 7.5. 7 Test methods 7.1 General Each individual lot not exceeding 100 000 units shall be subjected to testing, either individual or statistical. For statistical testing, the lot shall be homogenous and thermometers from various sources shall not be mixed. The sampling plan shall correspond to ISO 2859-2:1985, level II with a limiting quality level LQ = 5%. NOTE 1 Other sampling plans can be used if they are statistically equivalent. NOTE 2 For suggested types of testing, see Annex
SIST EN 12470-1:2000+A1:2009



EN 12470-1:2000+A1:2009 (E) 9 7.2 Method of test of the performance of pigment 7.2.1 General Testing of each batch shall be performed in accordance with 7.1. NOTE This requirement applies only to solid-stem thermometers. 7.2.2 Procedure Immerse the stem of the thermometer for 20 min in a solution of 700 g/l ethanol in water at 38 ûC and then wipe it dry. 7.3 Method of test for the determination of the average zero point depression 7.3.1 Construction of test thermometer Using the glass under consideration for the bulb, construct test thermometers with the following specifications: a) scale range at least: - 3,0 ûC to + 3,0 ûC; b) scale interval: 0,02 ûC, 0,05 ûC or 0,1 ûC; c) distance between consecutive scale lines: at least 1,0 mm; d) expansion chamber: of sufficient volume for the thermometer to be heated to 400 ûC without damage; e) stabilization: thermometers stabilized (see 7.3.2). 7.3.2 Confirmation of stabilization 7.3.2.1 Heat the thermometer in a suitable device e.g. a liquid bath or metal block oven, from ro
...

  • Relates Information
  • ISO 8130-9:1992

    ISO 8130-9:1992 - Coating powders
    09-28
  • EN 352-2:2020/FprA1

    EN 352-2:2021/oprA1:2023
    09-28
  • IEC TS 61158-4:1999

    IEC TS 61158-4:1999 - Digital data communications for measurement and control - Fieldbus for use in industrial control systems - Part 4: Data Link protocol specification Released:3/24/1999 Isbn:2831847656
    09-28
  • HD 566 S1:1990

    HD 566 S1:1998
    09-28
  • ISO 5131:1982/Amd 1:1992

    ISO 5131:1982/Amd 1:1992
    09-28
  • EN 60598-2-22:1990

    EN 60598-2-22:1996
    09-27
  • ISO 8504-2:1992

    ISO 8504-2:1992 - Preparation of steel substrates before application of paints and related products -- Surface preparation methods
    09-27
  • EN 12165:2024

    prEN 12165:2022
    09-27
  • IEC TS 61158-6:1999

    IEC TS 61158-6:1999 - Digital data communications for measurement and control - Fieldbus for use in industrial control systems - Part 6: Application Layer protocol specification Released:3/24/1999 Isbn:2831847613
    09-27
  • ISO 4252:1992

    ISO 4252:1992 - Agricultural tractors -- Operator's workplace, access and exit -- Dimensions
    09-27